X-ray radiation qualities

For calibrating diagnostic X-ray instruments, RRPPS produces the radiation qualities and irradiation conditions which are in accordance with the stated requirements of CEI/IEC 1267: 1994, Medical Diagnostic X-ray Equipment - Radiation Conditions for Use in the Determination of Characteristics.

A limited number of these beam qualities are included as standard; others are available by prior arrangement only.

Table 1 - Standard Radiation Qualities RQR: radiation qualities in radiation beams emerging from the X-ray tube assembly

Standard
radiation quality
Approximate
X-ray tube voltage
(kV)
Normal first
half-value layer
(mm Al)
RQR 2
40
1.0
RQR 3
50
1.5
RQR 4
60
2.0
RQR 5
70
2.5
RQR 6
80
2.9
RQR 7
90
3.3
RQR 8
100
3.7
RQR 9
120
4.5
RQR 10
150
5.7

Table 2 - Standard Radiation Qualities RQC: radiation quality in radiation beams emerging from the irradiated object simulating the patient

Standard
radiation quality
Approximate
X-ray tube voltage
(kV)
Normal first
half-value layer
(mm Al)
RQC
75
0.5

As part of our quality assurance and measurement traceability procedures, the radiation qualities specified by the IEC are cross referenced against the measured X-ray spectrum end point (peak photon energy), obtained using a hyper pure germanium detector.
 
Further details may be obtained from our paper "Development of a Calibration Facility for Test Instrumentation in Diagnostic Radiology", S.Green, J.E.Palethorpe, D.E.Peach & D.A.Bradley, published in Radiation Protection Dosimetry No 67, July 1996, pp41-46. For more information see external links.

Further information

For further information please contact:

  • Tracie Such (general enquiries)
  • Peter Austin (technical enquiries) 
  • Denise Delahunty (technical enquiries)

RRPPS
63 Melchett Road
Kings Norton Business Centre
Birmingham, B30 3HP
Telephone: 0121 627 2090/1
Fax: 0121 627 2057
Email: rrpps@uhb.nhs.uk

Links

External websites will open in a new browser window.

University Hospitals Birmingham NHS Foundation Trust is not responsible for the contents or the reliability of external websites and does not necessarily endorse the views expressed within them. Listing should not be taken as endorsement of any kind. We cannot guarantee that links to other websites will work all of the time, and we have no control over the availability of external web pages.

Getting here

Information about travelling to, staying at and getting around the hospital.

Getting to the hospital

Jobs at UHB

See why our hospitals are great places to work.

Jobs at UHB